I am using my HI 3010 Filler/Dispenser/IBC as a filler. I am using one of the inputs as an OK to DISCHARGE signal when I have a bag in place. The problem is that if the bag is not removed and replaced with a new bag, the OK to DISCHARGE signal is still present and the discharge will go into an already full bag. How can I set the unit up so it must see the bag as removed and replaced before discharging again?
Can I run the load cell cables in the same conduit as my 24 VDC digital signal wires and EtherNet cable? Does the milli-volt signal wire need to be shielded?
When I cycle power on my PLC, some of my HI 2151 Series units (HI 2151/20WC, HI 2151/30WC or the HI 2110WI) do not come online and show many retries, but others on the same network come online OK. When I change unit locations between one that does not come online with one that does, the problem stays with the location. What could be my problem?
Our country regulations require that if an indicator has lb / kg toggle key, it must be possible to lock, either by hardware or software. Is this possible with the HI 2110WI or HI 2151/30WC?
We are interfacing our HI 2204LT or HI 2201LT with an Allen-Bradley 1756-IF8 card. Can we use 24Vdc for the excitation voltage on a HI 2204LT or HI 2201LT? Would you have a wiring diagram for this?