Knowledgebase

Filter Filter button
Results: 21 - 30 of 78

Categories

Is the HI 215IT summing board compatible with all Hardy Instruments?

We are not getting a reading from the HI 215IT junction box. The HI 215IT is connected to a HI 1771-XX-RT and from there it is connected to a HI 1771WS module using the HI 1771-XX-C6 cable. Is the HI 215IT summing board compatible with all Hardy Instruments?

How can I tell if my load cell is good without using Integrated Technician?

I am using older load cells from Hardy Process Solutions (or from another company) and I want to make sure they are functioning properly before I use them in another application. I don't have the Hardy feature "Integrated Technician" in my junction box or controller, so I checked the installation/troubleshooting guide which says to check the resistance between the signal wires and the excitation wires and the resistance of each should be around 350 ohms. Well, the resistance between the signal wires on all 8 load cells was right at 350 ohms, but the resistance between the excitation wires on all 8 load cells was around 387 ohms. So, are all of these load cells bad, or is it ok for the excitation resistance to be 387 ohms?

Using a 4-Scale HI 3030, can I see individual load cell MV readings with IT?

Can I see individual load cell MV readings with IT on a HI 3030-4S (four scale option)?

What is the extra wire for on my Hardy Scale cable?

The cable coming out my ANYWEIGH bench or floor scale has an extra wire which is not called out in the manual. All load cell wires are identified, but this extra, heavier gauge wire is not. Should this wire be terminated anywhere or just capped off?

Can I mix different model load cells?

We have a system with 4 load cells and one load cell needs to be replaced. Do I have to use the same manufacturer and model, or can I mix different model load cells together on the same system? Will my C2 function still work?

Why can't we get a signal out of our summing junction box?

We cannot get a signal out of our summing junction box? What could be the problem?

Why is my C2® calibration incorrect?

Why is my C2® calibration incorrect?

Why are the relays clicking on my IT summing card?

 I am using a HI215IT-PS1 summing box, 3 Hardy load cells and a Hardy weight controller or PLC module. We are also using C2 cable between summing box and Hardy Instrument, and we are hearing a very loud clicking noise at the summing box. What can be causing this?

What is gravity correction and how do I use it?

What is gravity correction and how do I use it? 

How does C2® Calibration work with ISO 9000 certification?

How does C2® Calibration work with ISO 9000 certification?