Can I use NON C2® or NON IT Load Cells with Hardy Summing Boards or Junction Boxes?

I have an HI 215IT Hardy junction box with a label which reads: "CAUTION: USE ONLY WITH PRODUCTS HAVING IT CAPABILITIES". I've tried to connect different manufacturer load cells using this junction box, but I didn't read any values from the load cells. The load cells I have only have 4 wires, so they do not have C2 nor Integrated Technician (IT) capabilities. Can I use NON IT / NON C2 load cells with this junction box? I've read that Hardy controllers can work with and without C2 / IT load cells. Do I need a different junction box?

Why can't I see an exact zero reading?

When I send the command to zero or tare the scale the weight reading is not exactly zero. The weight reading goes to a value between +0.05 and -0.01 lbs, and the 3 load cells are 2000 lbs each. Why can't I see an exact zero reading?

Why are we seeing weighing errors using hard calibration?

We are trying to perform a hard calibration using 3000 lbs test weights. After Hard Cal our Hardy controller reads 3002 lbs with 3000 lbs applied, 1999.3lbs with 2000 lbs, and 995 lbs with 1000 lbs. At 0.0 the unit will not display a tenth of a pound without first applying about 8 lbs. When we suspend a test weight of 30 lbs on the cell the controller reads only 22 lbs. Why is there an error? When I used to calibrate a 4000 lb cell on our older HI 2151/20WC, we never saw more than a few tenths of a pound error throughout the test range 0-3000 lbs. Why the difference between the HI 2151/20WC and newer controllers?

How do I C2® calibrate my HI 2151/30WC or HI 2110WI?

How do I C2® calibrate my HI 2151/30WC or HI 2110WI?

What are the characteristics of the HI LPD load points?

What are the characteristics of the HI LPD load points?

Can my Remote I/O option board be swapped to a different Hardy Instrument?

Can my Remote I/O option board be swapped between the HI 2151/20WC, HI 2151/30WC, HI 2110WI, or HI 2160 rate controller?

Why isn't my weight reading stable after hard calibration?

After trying to do a new hard calibration on a previously working system, my weight reading is not stable and my Hardy Instrument will not calibrate properly. What could be the problem?

What are the HI 3010 or HI 3030 mounting options?

What mounting options are available for Hardy's HI 3010 or HI 3030?

Why can't I get my Hardy Instrument to zero?

I send a zero command to my Hardy Instrument and intermittently it will not zero. I can re-calibrate the unit and it will allow a zero command again for a while, then stop again. What can cause this?