Click Here to Learn More!If your HI2151/30WC unit is powered up, nothing appears on the main display, and the status bar LEDs are flashing on one side, this may indicate an issue with connectors between the internal circuit boards of your unit. In the past we have seen dust, corrosion, and/or other contaminants degrade the connection between the internal circuit boards.
The following steps may be tried to solve this issue:
NOTE: Be sure to de-energize the HI 2151/30WC before performing any other work. Use approved static control device to prevent damage to components when you are handling sensitive electronics!
Power up the unit. If the preceding recommendations do not rectify the problem, then the instrument should be returned to the factory for evaluation/repair. You can request a Return Authorization number online.