This procedure should only be used to restore known good traditional or hard calibration parameters on your HI 2151/30WC. If you used C2 calibration, just redo the C2 calibration. This is not a substitution for the calibration procedures outlined in the manual.
1. Prior to performing the following steps use previously recorded settings from self-test menus; the SPAN, Zr cnt, and FS cnt. The default setting in this hard cal submenu is not current and using incorrect settings will spoil your calibration. A common practice is to record these numbers after verifying the scale is weighing correctly. Then if the need arises you have a backup for a quick calibration recovery.
2. Check configuration entries for correct decimal position against the original parameters and enter the CALIBRATION menus until CAL is displayed.
3. (With CAL displayed)
4. Press: ENTER - Hd CAL now displayed
5. Press the MODE button (front display panel) - Zr cnt now displayed
6. Press ENTER - Displays current value
7. Press TST/CLR and Enter Zr cnt value
8. Press ENTER - SPAn displayed
9. Press ENTER - Displays current value
10. Press TST/CLR and Enter SPAN value
11. Press ENTER - FS cnt displayed
12. Press ENTER - Displays current value
13. Press TST/CLR and Enter Fs cnt value
14. Press ENTER - RETURN displayed
15. Press ENTER - ENDCAL displayed
16. Press ENTER - Current weight is displayed *
(*Automatically updates calibration into the Enhanced Secured Memory Module and exits Calibration)