You have no items in your shopping cart.

New HIDS Drum Scales!

Click Here to Learn More!
Support: 858-292-2710

How many load cells can I use with my HI 3030 or HI 3010?

Last updated 08/04/2014 03:26 AM

Question

How many load cells can a HI 3010 or HI 3030 power? Will C2® work with more than 4 load sensors?

Answer

The HI 3010 or HI 3030 single channel weight controllers will read and store calibration information for up to eight C2® load sensors and power up to eight 1000 ohm load cells, or four 350 ohm load cells.

A HI 3030-2S (2 channel) will power sixteen 1000 ohm load cells (max 8 per channel), or eight 350 ohm load cells. (max 4 per channel)

A HI 3030-4S (4 channel) will power thirty-two 1000 ohm load cells (max 8 per channel), or sixteen 350 ohm load cells (max 4 per channel), (16 load cells is the maximum current draw with the internal power supply). Additional load cells can be added, but will require an external 5vdc power supply.

Look up external power supply technote for additional details.