Click Here to Learn More!HI 3030 controller Parameter HELP menus:
Channel ID is a label to name a channel (Optional). You may enter up to 19 characters.
Instrument ID is a label (Optional). You may enter up to 19 characters.
Operator ID is a 3-character label identifying the operator (Optional).
Units of Measure may be lbs, Kgs, grams, or ounces. Weight is displayed in the chosen units. The units of measure may be changed without re-calibration.
Grad Size defines the incremental change between displayed weight readings.
Decimal Point defines the number of digits shown to the right of the decimal point.
Number of Averages is the number of weight reading updates averaged together to create the displayed weight. The unit updates at a rate of 55 times/sec. So a setting of 55 will averages 55 updates or 1 seconds worth of updates.
Motion Tolerance defines the amount the weight reading needs to change in a 1 sec period of time to make the scale go into MOTION status. If the change over the last 1 sec is less than the Motion Tolerance then the scale will not indicate in motion. Motion indicate means the weight on the scale is currently changing. The scale cannot be calibrated, tared, or zeroed while in motion.
Tare Weight is the difference between the gross and net weight.
Hardy Control-Link - The instrument transmits and receives data packets on Ethernet via UDP. The instrument will transmit packets at the rate selected if any destination is defined on its destination list (see the Hardy Control Link configuration page). This will set up your peer to peer communications.
Control-Link Port - This parameter determines the Ethernet port that packets are sent on. Note: the instrument listens and transmits on this port. A power cycle is required to change the listening port. Legal values are 1024-1055